Electronic And Mechanical Engineering Laboratory Report Accounting Essay

This standardization research lab involves the standardization of a burden cell and a spring balance utilizing a 10kg mass, adjusted by 1kg per measuring recorded giving 11 distinguishable readings of 0kg to 10kg in 1kg stairss. ANS/ISA recommend that this procedure should be repeated for a sum of 5 up-down rhythms to give a complete set of informations which would dwell of a sum of 110 readings, nevertheless due to clip restraints this lab will be restricted to 1 up-down rhythms giving a sum of 22 informations points. The indicated values at atmospheric force per unit area increasing were recorded along with the indicated values diminishing and these values were entered into a arrested development equation utilizing an excel spreadsheet. From this a graph was produced and a best fit curve tantrum to the measurings. We so set about finding the uncertainness bounds associated with the recession curve tantrum, the consequences of this showed that the burden cell had significantly less divergence from the true value over its full scope of values ; 1 standard divergence for the spring balance equated to 0.1238 while with the burden cell this was 0.00285. The burden cell was besides less susceptible to hysteresis with a full graduated table reading of 0.2 % whereas the spring balance had a full graduated table reading of 6.1 % , and besides non line rareness with a full graduated table reading of 0.2 % for the burden cell and 4.2 % for the spring balance.

Introduction

Calibration can be defined as a comparing between measurings, one of which has a known magnitude which has been set with a criterion with a known or assigned rightness and another measuring made in as similar a manner as possible with the unit under trial ( UUT ) or trial instrument ( TI ) . For any measuring to be considered accurate the indicated end product ( O ) of the measurement instrument must be every bit near as possible to the true value – which is besides known as the measureand ( I ) . This ‘closeness ‘ is quantified as a measuring mistake, as in the derived function between the indicated value and the true value.To guarantee the truth of equipment standardization is indispensable because the belongingss of constituents will necessarily float off from their set values over clip for many grounds such as emphasis on the constituent and environmental conditions, and the measurings made by these constituents become unsure overtime. Calibration establishes the uncertainness specifications that are caused by systematic and random mistakes associated with an instrument to be determined along with the inactive features of the instruments such as hysteresis and one-dimensionality.

We will write a custom essay sample on
Electronic And Mechanical Engineering Laboratory Report Accounting Essay
or any similar topic only for you
Order now

Test Equipment

Load Cell: A burden cell is a transducer which is used to change over a force into an electrical signal, it normally consists of 4 strain gages in a Wheatstone span formation. This is an indirect transition procedure, ab initio the force being sensed deforms the strain gage and so this distortion is converted to an electrical signal which is normally a few mVs and will necessitate to magnify before it can be utilized.

Jumping Balance: A spring balance is a weighing device that uses a normally additive relationship between a burden applied and spring distortion. The spring balance works on the rule of Hooke ‘s Law, this jurisprudence states that the force needed to widen a spring is relative to the distance that spring is extended from its remainder place and for this ground the graduated table markers on the spring graduated table are every bit spaced.

Method

This research lab was carried out based on the ANS/ISA ( 1979 ) a standard process for inactive standardization. The influence of modifying ( Im ) and interfering ( Ii ) influences were ignored, so the standardization was carried out under standard environmental conditions ( 200C and 1 ambiance ) where Im=Ii=0.

The standard weights used were so applied incrementally from Imax to Imin in 10 % increases giving 11 distinguishable readings on the up rhythm, so the weights were removed incrementally on the down rhythm supplying a farther 11 distinguishable readings. These readings were recorded from the spring balance graduated table in lbs, and besides from the transducer end product in mVs.

For truth ANS/ISA recommend that this procedure should be repeated for another 4 up-down rhythms to give a information set of 105 readings but for the intents of this research lab 22 information points were sufficient.

An excel spreadsheet was so used to suit a arrested development line to the information gathered. From here the true value, the systematic mistake and the random mistake for the burden cell were specified.

The associated European Norm criterions were so used to find if both instruments lie within the specified one-dimensionality and hysteresis features.

Consequences

Datas recorded:

Arrested development Analysis:

Using a arrested development analysis the standardization informations is fitted to a arrested development equation,

O = KI + a

Where O = Output or indicated value ( Dependent variable )

I = Input or true value ( Independent variable )

K = Slope of the line

a = Intercept on the perpendicular axis

K and a were so determined from arrested development analysis utilizing the undermentioned equation where N = figure of points:

For spring balance K = 22 ( 1689.2 ) – ( 110 ) ( 242.55 ) = 10481.9 = 2.165

22 ( 770 ) – 1102 4840

a = ( 242.55 ) ( 770 ) – ( 1689.2 ) ( 110 ) = 951.5 = 0.1965

22 ( 770 ) – 1102 4840

Spring balance arrested development equation from manual computation: O = 2.165I + 0.1965

Spring balance arrested development equation from excel: O = 2.1647I + 0.1994

Load cell arrested development equation from excel: O = 47.936I – 14.591

Post Calibration Application:

Post standardization in day-to-day usage, the burden cell will demo a value in millivolt and we need to be able to find about what the true value this will stand for in kilogram. To make this we rewrite the arrested development equation as a standardization equation as follows:

I = ( O – a ) /K = OK-1 – aK-1

Where I = an estimation of the true value utilizing the standardization equation based on the indicated value O

Spring balance standardization equation: I = O ( 0.4615 ) – 0.0920

Load cell standardization equation: I = O ( 0.0208 ) – 0.3034

Uncertainty Limits associated with Regression Curve Fit

The uncertainness associated with the arrested development curve tantrum ( O informations ) is given by:

Where: so = standard divergence associated with indicated values ( O )

The uncertainness associated with utilizing the reverse arrested development curve to foretell the true value on the footing of an indicated value is given by:

Depending on the degrees of uncertainness associated with the preciseness error the undermentioned specifications can be utilized:

0.674s ( 50 % assurance interval ) ( 50:100 )

s ( 69.3 % assurance interval ) ( 32:100 )

2s ( 95.4 % assurance interval ) ( 5:100 )

3s ( 99.7 % assurance interval ) ( 3:1000 )

Standard Deviation computation for Load Cell:

SUMMARY OUTPUT

Arrested development Statisticss

Multiple R

0.9999991

R Square

0.99999821

Adjusted R Square

0.99999812

Standard Error

0.21298741

Observations

22

Analysis of variance

A

df

United states secret service

Multiple sclerosis

F

Significance F

Arrested development

1

505536.9

505536.9

11144099.8

6.10719E-59

Residual

20

0.907273

0.045364

Entire

21

505537.8

A

A

A

A

Coefficients

Standard Error

T Stat

P-value

Lower 95 %

Upper 95 %

Lower 95.0 %

Upper 95.0 %

Intercept

-14.5909091

0.084953

-171.754

3.5923E-33

-14.768117

-14.4137

-14.7681

-14.4137

X Variable 1

47.9363636

0.01436

3338.278

6.1072E-59

47.90641003

47.96632

47.90641

47.96632

RESIDUAL OUTPUT

Observation

Predicted Yttrium

Remainders

1

-14.5909091

0.090909

0.00826446

2

33.3454545

-0.04545

0.00206612

3

81.2818182

0.018182

0.00033058

4

129.218182

-0.01818

0.00033058

5

177.154545

0.045455

0.00206612

6

225.090909

-0.09091

0.00826446

7

273.027273

-0.02727

0.0007438

8

320.963636

0.036364

0.00132231

9

368.9

0.1

0.01

10

416.836364

-0.83636

0.69950413

11

464.772727

0.227273

0.05165289

12

464.772727

0.227273

0.05165289

13

416.836364

0.163636

0.02677686

14

368.9

0.1

0.01

15

320.963636

0.036364

0.00132231

16

273.027273

-0.02727

0.0007438

17

225.090909

-0.09091

0.00826446

18

177.154545

0.045455

0.00206612

19

129.218182

0.081818

0.00669421

20

81.2818182

0.018182

0.00033058

21

33.3454545

0.054545

0.00297521

22

-14.5909091

-0.10909

0.01190083

amount

0.90727273

So sqd

0.04123967

So

0.20307553

Si

0.00423636

Calibration Equation for Load Cell:

0.674s ( 50 % assurance interval ) ( 50:100 ) = 0.00285

s ( 69.3 % assurance interval ) ( 32:100 ) = 0.00423636

2s ( 95.4 % assurance interval ) ( 5:100 ) = 0.008472

3s ( 99.7 % assurance interval ) ( 3:1000 ) = 0.0127

Standard Deviation for Spring Balance:

SUMMARY OUTPUT

Arrested development Statisticss

Multiple R

0.998318

R Square

0.996638

Adjusted R Square

0.99647

Standard Error

0.41699

Observations

22

Analysis of variance

A

df

United states secret service

Multiple sclerosis

F

Significance F

Arrested development

1

1030.865

1030.865

5928.58

3.26E-26

Residual

20

3.477611

0.173881

Entire

21

1034.342

A

A

A

A

Coefficients

Standard Error

T Stat

P-value

Lower 95 %

Upper 95 %

Lower 95.0 %

Upper 95.0 %

Intercept

0.199432

0.166321

1.199075

0.244513

-0.14751

0.546372

-0.14751

0.54637205

X Variable 1

2.164659

0.028113

76.99728

3.26E-26

2.106015

2.223303

2.106015

2.223302693

RESIDUAL OUTPUT

Observation

Predicted Yttrium

Remainders

1

0.199432

-0.07443

0.00554

2

2.364091

-0.23909

0.057164

3

4.52875

-0.27875

0.077702

4

6.693409

-0.19341

0.037407

5

8.858068

-0.10807

0.011679

6

11.02273

-0.22273

0.049607

7

13.18739

-0.18739

0.035114

8

15.35205

-0.15205

0.023118

9

17.5167

-0.2667

0.071131

10

19.68136

-0.38136

0.145438

11

21.84602

-0.64602

0.417345

12

21.84602

-0.64602

0.417345

13

19.68136

0.918636

0.843893

14

17.5167

0.733295

0.537722

15

15.35205

0.447955

0.200663

16

13.18739

0.312614

0.097727

17

11.02273

0.377273

0.142335

18

8.858068

0.391932

0.153611

19

6.693409

0.306591

0.093998

20

4.52875

-0.02875

0.000827

21

2.364091

0.135909

0.018471

22

0.199432

-0.19943

0.039773

amount

3.477611

So sqd

0.158073

So

0.397584

Si

0.183671

Calibration Equation for Spring Balance:

0.674s ( 50 % assurance interval ) ( 50:100 ) = 0.1238

s ( 69.3 % assurance interval ) ( 32:100 ) = 0.18371

2s ( 95.4 % assurance interval ) ( 5:100 ) = 0.36742

3s ( 99.7 % assurance interval ) ( 3:1000 ) = 0.55113

Bias and Precision Error of the Instrument:

The procedure of standardization allows the entire mistake to be separated into two separate parts, the prejudice mistake and the impreciseness mistake. For illustration for a Load Cell reading of 230mV, the best estimation of true value is 4.85kg, the prejudice is observed as -0.3034 giving an indifferent best estimation of 4.5kg and the impreciseness associated with this information is given by A± 0.0127 ( 3s ) bounds.

One-dimensionality

A measurement instrument is determined to be additive if all the corresponding values of I and O are in a consecutive line over the scope of the instrument if this is non the instance the instrument is so said to be non additive. Most instruments would hold some non one-dimensionality.

Non-Linearity

Non-Linearity ( % FS )

Scale

Load Cell

Scale

Load Cell

0.13

0.09

0.6 %

0.0 %

0.24

0.07

1.1 %

0.0 %

0.28

0.03

1.3 %

0.0 %

0.20

0.10

0.9 %

0.0 %

0.12

0.06

0.6 %

0.0 %

0.23

0.22

1.1 %

0.0 %

0.20

0.19

0.9 %

0.0 %

0.17

0.15

0.8 %

0.0 %

0.29

0.11

1.3 %

0.0 %

0.40

1.08

1.9 %

0.2 %

0.67

0.04

3.2 %

0.0 %

0.67

0.04

3.2 %

0.0 %

0.90

0.08

4.2 %

0.0 %

0.71

0.11

3.4 %

0.0 %

0.43

0.15

2.0 %

0.0 %

0.30

0.19

1.4 %

0.0 %

0.37

0.22

1.7 %

0.0 %

0.38

0.06

1.8 %

0.0 %

0.30

0.00

1.4 %

0.0 %

0.03

0.03

0.2 %

0.0 %

0.13

0.03

0.6 %

0.0 %

0.20

0.11

0.9 %

0.0 %

0.90

1.08

4.2 %

0.2 %

Maximal non-linearity of Spring Balance = 0.90 = 4.2 % Full graduated table reading

Maximal non-linearity of Load Cell = 1.08 = 0.2 % Full graduated table reading

Hysterises

Hysteresis is described as for a given input, the end product displayed by the instrument may be different depending if the value is increasing or diminishing. The hysterises measuring is the difference of the end product measuring for the same input between the up rhythm and the down rhythm.

Maximal hysterises of Spring Balance = 1.3 = 6.1 % Full graduated table reading

Maximal hysterises of Load Cell = 1.0 = 0.2 % Full graduated table reading

Discussion and Decision

Calibration is indispensable to guarantee the truth of measurement instrument and this research lab set out to show the methods used to find the truth degrees of a burden cell and a spring balance. This occurred under standard environmental conditions i.e. the modifying and meddlesome influences were ignored. Quite frequently the end product from a measuring instrument could be affected by factors other than the input, such as atmospheric force per unit area, ambient temperature, supply electromotive force or comparative humidness Standard environmental conditions are defined as 25oC ambient temperature, 1 saloon atmospheric force per unit area, 50 & A ; comparative humidness and 10V supply electromotive force. There are 2 types of environmental inputs, modifying that can do the additive sensitiveness of a measurement instrument to alter most commonly temperature and interfering that can do the intercept to alter from standard value largely caused by electrical intervention. We chose to breathe these factors due to the stable temperature in the room and the fact that there was virtually no possibility of electrical intervention.

The truth of the burden spring readings could be improved greatly if the graduated table had a better declaration besides at that place would hold been important mistakes introduced to the informations by virtuousness of the fact that the criterions were non in 10 % graduations. The burden and droping between each reading most would hold destroyed the truth of the hysteresis informations gathered. Besides if the recommended 105 readings had been gathered a more accurate set of informations could be produced.

The standardization equation showed that the burden cell was much more accurate that the spring balance with a 99.7 % assurance interval with a standard divergence of merely 0.0127 for the burden cell as opposed to 0.55113 for the spring balace.There is much less hysteresis on the burden cell with merely 0.2 % of full graduated table reading as opposed to 4.2 % for the spring balance.

×

Hi there, would you like to get such a paper? How about receiving a customized one? Check it out